RF MIC2 AC Tag
Product description
Especially made to be Autoclave resistant
Chemical sterilization for decontamination
Liquid chemical sterilants include glutaraldehyde, hydrogen peroxide, hydrogen peroxide-peracetic acid combinations, sodium hypochlorite, and peracetic acid. Pass
Thermal (hot water) disinfection
- Washer-sanitizers, washer-decontaminators, utensil washers, and cart washers reduce microbial contamination by means of cleaning agents, hot water, rinsing, and drying.
- Pasteurization equipment provides cleaning and high-level disinfection at water temperatures of 65°C to 77°C (150°F to 170°F) for a contact time of at least 30 minutes.
- Washer-disinfectors provide a cycle of cleaning, rinsing, disinfection, and drying at temperatures that are usually higher than those of washer-sanitizers. Pass
Thermal sterilization for decontamination
Saturated steam can be used to decontaminate devices capable of withstanding high temperatures (121°C to 135°C [250°F to 275°F]) and pressures (16 to 35 psig). Washer-sterilizers provide cleaning and rinsing followed by exposure to saturated steam at temperatures of 121°C to 135°C (250°F to 275°F). Pass
Ultrasonic cleaner
30 KHz for 30 min Pass
Dry heat sterilization
160°C for 2 h or 170°C for l h or 180°C for 30 min Pass
Disinfection by light
C wave ultraviolet and wavelength range is from 200 nrn to 275nrn Pass
Detergent enzyme cleaner Pass
Autoclave sterilization
121°C and 115 Kpa for 1000 cycles Pass